- Introduction
- Features
- Specifications
Total phosphate refers to the sum of orthophosphate, pyrophosphate, metaphosphate, polyphosphate and organic phosphate in water. It is the result of the determination of various forms of phosphorus converted into orthophosphate after high temperature digestion of water sample.
The monitor has the functions of controlling the water sample collection automatically, high temperature digestion, reaction measurement, data processing and data transmission, which can finish the whole process of total phosphate analysis of wastewater fast, continuously and automatically.
According to the ammonium molybdate spectrophotometry, in the acid solution, the decomposer, potassium persulfate, converts polyphosphate and organic phosphorus into orthophosphate. The orthophosphate reacts with ammonium molybdate to form a yellow phosphomolybdate antimony complex, the complex is reduced to phosphomolybdate blue by ascorbic acid, and then the content of phosphate is determined by the total phosphate monitor.
◆ The integrated design of digestion measurement makes the measurement result more accurate and faster.
◆ Imported monochromatic light source – Excellent performance, concentrated energy, and low power consumption.
◆ High precision injection pump – High measurement accuracy, low reagent consumption.
◆ Acid-resistant and alkali-resistant, high temperature resistant valves -- Long service life and less manual maintenance.
◆ Equipped with pretreatment device, which can meet the measurement requirements of different water quality.
◆Auto-calibration function -- ensures more accurate measurement.
◆ With intelligent maintenance function -- It can clean automatically after each detection, which ensures more accurate measurement each time and reduces the manual maintenance.
Measurement Method |
Ammonium molybdate spectrophotometry |
Measurement Range |
0-1.5mg/L; 0-15mg/L; 0-50mg/L. The range can be extended. |
Lower Limit of Detection |
0.05mg/L |
Resolution |
0.01mg/L |
Display error |
0-1.5mg/L: ±0.05mg/L; 0-15mg/L: ±8%; 0-50mg/L: ±10% |
Repeatability |
<5% |
Span Drift |
No more than ±5%FS within 24h |
Measurement Interval |
1h-24h, optional |
Measurement Type |
Cyclic measurement, Fixed-point measurement, Trigger measurement |
Ambient Temperature |
5-45℃ |
Ambient Humidity |
<95% |
Display |
7” LCD touch screen |
Language |
Chinese / English |
Output |
4-20mA |
Communication |
RS485,MODBUS |
Power Supply |
AC(100-240)V; 50HZ |
Power Consumption |
<100W |
Overall Dimension |
500mm*395mm*1450mm (L*W*H) |
Weight |
About 60kg |
Other Functions |