- Introduction
- Features
- Specifications
Applications:
. Power. Chem or Petrochemical
. Chemical Fibre
. Panel mount,aluminum enclosure
---anti-interference
. 192×64 lattice LCD display
. Diaphragm-type precise metering pump
---anticorrosion,accurate and reliaable measurement
---maintenance free
. 460nm light source
---long life,stable and reliable
. Unique blank compensation technology
---drift free and accurate measurement
. Programmable multi-stream switching up to 6 streams
. Freely set between 10~60Min
. Large datalog of previous measurements and trend curve trackable
Display |
320×240 lattice LCD, Chinese or English |
Measurement Range |
(0-100)μg/L |
Measurement Cycle |
freely set between 5~60 min. |
Response Time (T90) |
5 min at 25oC |
Accuracy |
±2%F.S |
Repeatability |
±1% |
Stability |
Baseline drift::it can not be detected by the blank calibration Chemical drift: ±1%F.S/24h (depend on reagent stability) |
Sample Condition |
Temperature: 5 to 50oC Flow: (10-40)L/h Solid contents in sample: ≤5μm (no gel and grease) |
Ambient Temperature |
5 to 45oC |
Ambient Humidity |
≤90%RH (no condensation) |
Reagent Number |
One |
Reagent Consumption |
3L/month |
Analog Output |
(0-10)mA, (0-20)mA or (4-20)mA (each channel has its isolated output) |
Power Supply |
(220±22) VAC, (50±1) Hz |
Power Consumption |
≤150W |
Overall Dimension (mm) |
690(L) × 450(W) × 215(D) |
Cutout Dimension (mm) |
645 × 410 |
Net Weight: |
22kg |
Alarm Outputs |
No flow, high alarm and low alarm (each channel output individually) |
Hydrazine standard solution
1bottle
Drain pipe
2m
Φ8mm*2 tube
1piece
Spanner
1piece
Screw driver
1piece
Key
1piece
Reagent tube
1piece
Organic glass filter of each stream
1set
Connector of each stream
1set
Flow water tank kits of each stream
1set
Φ6 plastic 3-way connector of each stream
1piece