HK-368 Inductive Concentration/Conductivity Analyzer is an online industrial analyzer. The sensor adopts corrosion-resistant material and contact-free detection mode, which can effectively prevent the sensor from being corroded, polluted and polarized effect by the corrosive medium. There are many installation method, which can be suitable for different occasions. The measuring medium can be switched, which is easy to use. It is mainly used to monitor the regenerated solution concentration of ion exchanger in the process of producing high-purity water by ion exchange. It can be widely used in power, chemical industry and seawater to continuously measure the low-concentration acid/alkali/salt.
◆ Adopts corrosion-resistant materials, which can prevent thepolarization, scaling problems.
◆ Therange of conductivity is0—2000 mS/cm, the sensor does not need to be replaced.
◆ It can measure conductivity,concentration (NaOH, NaCl, HCl andH2SO4), salinity and TDS.
◆ Chinese or English menu, and easy to set the parameters.
◆ The color LCD screen can display the measured value, sample temperature and measuring medium, the content is rich and easy to understand.
◆ 100,000 measurement data can be saved, which is convenient to view.
◆ Two current outputs (4-20mA) and four alarm signals.
◆ Communication:RS485/PROFIBUS-DP/HART, choose one.
◆ Flange installation method or flow cell installation method is selected for different site environment.
Display: |
Chinese/ English, TFT color LCD |
|||
Measurement Range: |
Conductivity: NaOH : HCl: NaCl: H2SO4: HNO3: Salinity: TDS: |
0~200ms/cm; 0~2000ms/cm; 0~10.00%; 0~10.00%; 0~25.00%; 0~10.00%; 0~10.00%; 0~80.00ppt; 0~80.00g/L |
||
Measuring Medium: |
NaOH, HCl, NaCl, H2SO4, HNO3 |
|||
Resolution: |
Conductivity: 1mS/cm; Concentration: 0.01% |
|||
Error: |
Conductivity: 2% FS; Concentration: ±5%FS |
|||
Repeatability: |
±0.5FS% |
|||
Stability: |
±1.0%FS/30min |
|||
Temperature Sensor: |
PT1000 |
|||
Temperature Accuracy: |
±1℃ |
|||
Temperature Solution: |
0.1℃ |
|||
Temperature Compensation Range: |
(0~55)℃ |
|||
Current Output: |
(0~10)mA, (0~20)mA or (4~20)mA (Dual-channel independent isolated output), optional. Load: 600Ω, Error: 0.1mA |
|||
Communication: |
RS485, Profibus or Hart (optional) |
|||
Relay Output: |
Dual-channel configurable |
|||
Storage: |
≥100,000 |
|||
Response time: |
≤30s |
|||
Sample Temperature: |
(0~40)℃ |
|||
Working Pressure of Flow Cell Sensor: |
≤0.25MPa |
|||
Medium Flow of Flow Cell Sensor: |
(6~20)L/h |
|||
Solid Contents in Sample: |
≤5μm (no gel and grease) |
|||
Power Supply: |
AC(100~240)V, (50~60)Hz |
|||
Ambient Temperature: |
||||
Ambient Humidity: |
≤90%RH (No condensation) |
|||
Storage temperature: -20℃~55℃ |
||||
Safety Requirement |
Leakage Current |
≤3.5mA |
||
Ground Resistance |
≤0.5Ω |
|||
Insulation Resistance |
The insulation resistance of the power phase and the neutral line of the electronic unit to the ground ≥ 40MΩ. |
|||
Dielectric Strength |
Between the power phase, neutral line and ground, and between the power input terminal and the metal parts that may be touched, it should be able to withstand a test voltage of AC 1500V, 50Hz for 1min, and there should be no electric breakdown or arcing. |
|||
Static Test |
For the air discharge test, apply 8kV voltage to the panel, shell, cable, and interval 1min. For the contact discharge test, apply 6kV voltage to the panel, shell, cable, and interval 1min. There should be no crashes. |
|||
Transmitter Overall Dimension:
|
146mm*146mm |
|||
Sensor Overall Dimension: |
146mm*146mm*146mm |
|||
Sensor Cable Length: |
5m (Special requirements can be customized before ordering) |
|||
Keypad: |
8 keys |
|||
Installation Method: |
Panel-mounting, Wall-mounting, Pipe-mounting |
|||
Power Consumption: |
≤8W |
|||
Weight: |
5kg |
Transmitter: 1 set
Sensor: 1 set
Installation component: 1 set (choose one)
Choose one component according to the user’s order |
|
Flow cell: 1 set |
Flange: 1 set |
|
|