- Introduction
- Features
- Specifications
. SMT and integrative design
. Rugged cast aluminum enclosure
---anti-static,anti-interference,easy-to-install
. 192×64 lattice LCD display
. User interface language:English
. Simultancous display of multtiparameters
. Automatic or manual temperature compensation
. Unique 4-wire measuring technology
---neglect the effect made on measurement by cable length
. Two analog outputs and two alarm relay outputs
. CAN bus communication
. Large datalog and history curve trackable
. Universal power supply for worldwide operation(100~240)V
Display |
192×64 lattice LCD |
Measurement Range |
K=0.01 (0.000-2.000)μS/cm, (0.00-20.00)μS/cm K=0.10 (0.00-20.00)μS/cm, (0.0-200.0)μS/cm K=1.00 (0.0-200.0)μS/cm, (0-2000)μS/cm K=10.0 (0.000-2.000)mS/cm, (0.00-20.00)mS/cm |
Temperature Range |
(0.0 to 99.9)oC |
Temperature Compensation |
(0.1 to 60.0)oC (auto) |
Resolution (min) |
Conductivity: 0.001μS/cm Temperature: 0.1oC |
Accuracy |
Conductivity: ±1%F.S; Temperature: ±0.5 oC |
Response Time (T90) |
30 sec at 25oC |
Analog Output |
(0~10), (0~20) or ( 4~20)mA Two (0~10)mA, (0~20)mA or (4~20)Maisolated outputs max load 800W(Correspond to conductivity and temperature) |
Output Accuracy |
±1%F.S |
Alarm Output |
Two alarm relays, programmable |
Communication |
CAN Bus (option) |
Sample Condition |
Temperature: (5~50) oC Flux :≤6L/H Solid contents in sample: ≤5mm (no gel and grease) |
Ambient Temperature |
(5 to 45)oC |
Ambient Humidity |
≤90%RH (no condensation) |
Storage Temperature |
(-25 to 55)oC |
Overall Dimension (mm) |
158(W) × 158(H) × 170(D) |
Cutout Dimensions (mm) |
140 × 140 |
Power Supply |
(100-240)VAC, (50-60)Hz |
Power Consumption |
≤8W |
Insulation Strength |
Power to ground 1500V, no flying arcin 1 min |
Weight |
2kg |
Analyzer
1piece
Conductivity cell
1piece
Cable
10m or 20m optional
Flow cell kits
1set
Simple calibrator
According to different kinds of installation
PG9 connector
4pcs
Cross head screw
2pcs