fault |
Analysis |
Repair method |
LCD blurred screen, no display,
|
1、LCD damaged |
Change LCD |
2、LCD cable connect wrong |
Connect the LCD again |
|
3、Power board and main board contact damaged. |
Reconnect the connection wire |
|
7、Main board has some problem |
Change the |
|
Unstable measurement, slow reaction or instument cannot calibrated normally |
1. Water sample has the different pressure |
Change water flow, keep stable |
2. Electrode is polluted |
Clean the electrode, immerse the electrode in the 3mol/L KCL solution for 24 hours |
|
3. High resistant has problem |
Check and change the cable if necessaty |
|
4. Electrode exceed time limit or damaged |
Change electrode |
|
5. Electric equipment faults |
Return back to factory for repair |
|
Test data shows maximum or minimum or no change |
1. Bad drainge, there is pressure at water export, which cause test result no change |
1. Try to make the water drain directly. |
2. Flow cell grounding cause the drift to the max or min value |
Change the location of flow cell to solve the problem |
|
3. Power grouding badly |
Connect to the valid power grouding |
|
4. The temp electrode insulation strength is not enough. |
Change temp electrode or remove temp electrode, or change it to manual temp compensation |
|
5. Instrument calibration is wrong. |
Do the calibration in the right way. |